M3A Vital Signs Monitor User Manual Specifications
- 92 -
Neonatal Mode
Default: 100 mmHg
Range: 60/70/80/100/120 mmHg
A1.2.9 SpO
2
Complies with ISO 80601-2-61: 2011.
Measuring Range
0% ~100%
Adjustable Range of Alarm Limits
0% ~100%
Resolution
1%
Data update period
1s
Accuracy
ADU & PED
2% (70% - 100% SpO
2
)
Undefined (0% - 69% SpO
2
)
NEO
3% (70% - 100% SpO
2
)
Undefined (0% - 69% SpO
2
)
Perfusion Index
Measuring Range
0-10. It displays 0 for invalid PI value.
Resolution
1 bpm
Sensor
Red light
(6603) nm
Infrared light
(90510) nm
Emitted light energy
< 15 mW
NOTE:
The information about wavelength range can be especially useful to clinicians (for
instance, when photodynamic therapy is performed).
A1.2.10 PR
Measuring range
Accuracy
Resolution
PR (SpO
2
)
25 bpm to 300 bpm
2 bpm
1 bpm
PR (NIBP)
40 bpm to 240 bpm
±3 bpm or 3.5%, whichever is greater
1 bpm

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN M3A and is the answer not in the manual?

EDAN M3A Specifications

General IconGeneral
TypeVital Signs Monitor
SpO2 Measurement Range0-100%
ParametersSpO2, Pulse Rate, Temperature
Pulse Rate Measurement Range30 to 250 bpm
Temperature Measurement Range0 to 50°C
Power SupplyAC 100-240V, 50/60Hz
Battery LifeUp to 8 hours

Related product manuals