M3A Vital Signs Monitor User Manual Specifications
- 91 -
Measuring range
ADU mode
SYS 40 mmHg ~270 mmHg
DIA10 mmHg ~ 215 mmHg
MAP 20 mmHg ~235 mmHg
PED mode
SYS 40 mmHg ~230 mmHg
DIA 10 mmHg ~180 mmHg
MAP 20 mmHg ~195 mmHg
NEO mode
SYS 40 mmHg ~ 135 mmHg
DIA 10 mmHg ~100 mmHg
MAP 20 mmHg ~110 mmHg
Alarm type
SYS, DIA, MAP, PR
Cuff Pressure measuring range
0 mmHg ~300 mmHg
Pressure resolution
1 mmHg
Maximum mean error
5 mmHg
Maximum standard deviation
8 mmHg
Maximum measuring time of single
measurement
ADU/PED 120 s
NEO 90 s
Typical measuring period
20 s ~ 35 s (depend on HR/motion disturbance)
Overpressure protection
ADU
(297±3 ) mmHg
PED
(245±3 ) mmHg
NEO
(147±3 ) mmHg
PR
Measuring range
40 bpm ~ 240 bpm
Accuracy
±3 bpm or 3.5%, whichever is greater
Pre-inflation Pressure
Adult Mode
Default: 160 mmHg
Range: 80/100/120/140/150/160/180/200/220/240
mmHg
Pediatric Mode
Default: 140 mmHg
Range: 80/100/120/140/150/160/180/200 mmHg

Table of Contents