EasyManuals Logo

EDAN M3A User Manual

EDAN M3A
111 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #98 background imageLoading...
Page #98 background image
M3A Vital Signs Monitor User Manual Specifications
- 91 -
Measuring range
ADU mode
SYS 40 mmHg ~270 mmHg
DIA10 mmHg ~ 215 mmHg
MAP 20 mmHg ~235 mmHg
PED mode
SYS 40 mmHg ~230 mmHg
DIA 10 mmHg ~180 mmHg
MAP 20 mmHg ~195 mmHg
NEO mode
SYS 40 mmHg ~ 135 mmHg
DIA 10 mmHg ~100 mmHg
MAP 20 mmHg ~110 mmHg
Alarm type
SYS, DIA, MAP, PR
Cuff Pressure measuring range
0 mmHg ~300 mmHg
Pressure resolution
1 mmHg
Maximum mean error
5 mmHg
Maximum standard deviation
8 mmHg
Maximum measuring time of single
measurement
ADU/PED 120 s
NEO 90 s
Typical measuring period
20 s ~ 35 s (depend on HR/motion disturbance)
Overpressure protection
ADU
(297±3 ) mmHg
PED
(245±3 ) mmHg
NEO
(147±3 ) mmHg
PR
Measuring range
40 bpm ~ 240 bpm
Accuracy
±3 bpm or 3.5%, whichever is greater
Pre-inflation Pressure
Adult Mode
Default: 160 mmHg
Range: 80/100/120/140/150/160/180/200/220/240
mmHg
Pediatric Mode
Default: 140 mmHg
Range: 80/100/120/140/150/160/180/200 mmHg

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN M3A and is the answer not in the manual?

EDAN M3A Specifications

General IconGeneral
TypeVital Signs Monitor
SpO2 Measurement Range0-100%
ParametersSpO2, Pulse Rate, Temperature
Pulse Rate Measurement Range30 to 250 bpm
Temperature Measurement Range0 to 50°C
Power SupplyAC 100-240V, 50/60Hz
Battery LifeUp to 8 hours

Related product manuals