M3A Vital Signs Monitor User Manual TEMP Monitoring (Optional)
- 72 -
2 The device must stay in stable ambient (room) temperature for 30 minutes before
operation.
3 Before the measurement, please stay in a stable environment for five minutes and
avoid exercise or bath for 30 minutes.
4 It is recommended that you measure the same ear for three times. If the three
measurements are different, select the highiest temperature.
5 Remember to compare the measurement result to the regular temperature of the
patient.
6 There is no gender and age limitation for using infrared ear thermometer.
7 The data saved in the thermometer is the last measurement data before the
thermometer is powered off.
8 Clinical repeatability: 0.12 C (1~5 years old); 0.10 C (> 5 years old).
10.2.3 TEMP Setup
Click on the TEMP Setup in the Main Menu:
TEMP Unit: Set temperature unit to C or ℉.
10.2.4 Alarm
10.2.4.1 Adjustable Range of Alarm Limits
The alarm limits are as follows:
+35.5 C (+95.9 ℉)~+42 C (+107.6 ℉)
10.2.4.2 Alarm Information and Prompts
Physiological alarms:
Measuring value of TEMP is above upper alarm limit.
Measuring value of TEMP is below lower alarm limit.