M3A Vital Signs Monitor User Manual TEMP Monitoring (Optional)
- 66 -
WARNING
9 Use disposable TEMP sensor covers recommended by EDAN to limit patient
cross-contamination. The use of any other probe cover may produce temperature
measurement errors or result in inaccurate readings.
10 Make sure you have set the correct measuring position before you start a
measurement, or the result will be inaccurate.
11 Do not move the probe during measurement.
12 In monitoring mode, no physiological alarms are available.
NOTE:
The cumulative use time for the oral or rectal temperature probe in a single patient should
be less than 24 hours.
10.1.2 Measuring Procedure
1 Ensure the sensor are well installed. There are icons indicating TEMP measuring position on
the main interface. If changing measuring position or measuring mode is necessary, enter menu
for setting.
2 Take out the sensor from the sensor bracket. After warm-up, it beeps and displays prompt for
starting TEMP measuring in information area.
3 Load a sensor cover by inserting the sensor into a sensor cover and press the sensor handle
firmly. The sensor handle will move slightly to engage the sensor cover.
4 Holding the sensor handle with your thumb and two fingers, and insert it to the measuring
position.
For measuring oral TEMP, place the sensor tip under the patients tongue on either side of the
mouth to reach the rear sublingual pocket. Have the patient close his lips around the sensor.
Figure 10-1 Measuring Position in Mouth
For measuring oral TEMP, do not take an axillary temperature through patient’s clothing.
5 The monitor enters Predict measuring mode, displays in the TEMP parameter area.
Sublingual Pocket

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the EDAN M3A and is the answer not in the manual?

EDAN M3A Specifications

General IconGeneral
TypeVital Signs Monitor
SpO2 Measurement Range0-100%
ParametersSpO2, Pulse Rate, Temperature
Pulse Rate Measurement Range30 to 250 bpm
Temperature Measurement Range0 to 50°C
Power SupplyAC 100-240V, 50/60Hz
Battery LifeUp to 8 hours

Related product manuals