M3A Vital Signs Monitor User Manual TEMP Monitoring (Optional)
- 80 -
Technical alarms:
TEMP module
failure or
communication
failure.
Stop using measuring function
of TEMP module; notify
biomedical engineer or
Manufacturer’s service staff.
The TEMP value is
out of the range of
+30 C ~ +43 C.
Put the probe into the probe
well; take it out and measure
again.
Probe configuration
(or no probe
connected) error.
Well connect the probe and the
monitor, and measure again.
Incorrect sensor type
or incorrect measure
sites
Use the correct type of sensor;
choose the correct measesure
site.
Unable to pre-heat
probe tip.
Stop using measuring function
of TEMP module; notify
biomedical engineer or
Manufacturer’s service staff.
NOTE: Measure readings
displayed on the screen are
unreliable when the monitor
indicates Temp Error P06.
System error during
synchronization.
Stop using measuring function
of TEMP module; notify
biomedical engineer or
Manufacturer’s service staff.
System error during
patient thermistor
calibration.
System error during
heater thermistor
calibration.
Monitor Mode patient
thermistor unstable or