M3A Vital Signs Monitor User Manual TEMP Monitoring (Optional)
- 80 -
Technical alarms:
Message
Cause
Alarm Level
What to do
Temp COMM
STOP
TEMP module
failure or
communication
failure.
High
Stop using measuring function
of TEMP module; notify
biomedical engineer or
Manufacturer’s service staff.
Temp exceed
limit
The TEMP value is
out of the range of
+30 C ~ +43 C.
Medium
Put the probe into the probe
well; take it out and measure
again.
No TEMP
SENSOR
Probe configuration
(or no probe
connected) error.
Low
Well connect the probe and the
monitor, and measure again.
TEMP SENSOR
ERR
Incorrect sensor type
or incorrect measure
sites
Medium
Use the correct type of sensor;
choose the correct measesure
site.
Temp Error P06
Unable to pre-heat
probe tip.
Low
Stop using measuring function
of TEMP module; notify
biomedical engineer or
Manufacturer’s service staff.
NOTE: Measure readings
displayed on the screen are
unreliable when the monitor
indicates Temp Error P06.
Temp Error E01
System error during
synchronization.
Medium
Stop using measuring function
of TEMP module; notify
biomedical engineer or
Manufacturer’s service staff.
Temp Error E02
System error during
patient thermistor
calibration.
Medium
Temp Error E03
System error during
heater thermistor
calibration.
Medium
Temp Error E04
System timing error.
Medium
Heater error
Heater error.
Medium
Temp Error P02
Monitor Mode patient
thermistor unstable or
Low