2.4.2 Functional Specifi cations
Function / Model InfusoVet Connect
Infusion Mode
ml/h Mode ●
Body-weight mode ●
Drip mode ●
Drug library mode ○
Micro Mode ●
Sequence Mode ●
Occlusion alarm level
5 Levels adjustable:
Level 1: 50mmHg
Level 2: 150mmHg
Level 3: 300mmHg
Level 4: 600mmHg
Level 5: 900mmHg
Drug library ≥2,000
History entries ≥5,000
Brand Library ≥200
WiFi ○
Remarks: ● means standard; ○ means optional.
Attention!
This User Manual describes the most confi guration and most complete functions, due to model difference or
optional components, not all functions are equipped in the product you purchased.
2.5 Product Specifi cation
Safety Classifi cation
Electric protection Type Class I
Electric protection Level Defi brillation proof type CF applied Part
Protection against fl uid
ingress
IP44
Working mode Continuous operation
Classifi cation Portable equipment, non-portable infusion pump
Specifi cation Parameters
Infusion apparatus
specifi cation
10 – 60 drops/ml
Infusion Rate
10 – 20 drops/ml specifi cation infusion apparatus: 0.1 – 2,000ml/h
21 – 40 drops/ml specifi cation infusion apparatus: 0.1 – 800 ml/h
41 – 60 drops/ml specifi cation infusion apparatus: 0.1 – 400 ml/h
Minimum step is 0.01 ml/h
< 100 ml/h step is 0.01ml/h
< 1,000 ml/h step is 0.1ml/h
≥ 1,000 ml/h step is 1ml/h