USER MANUAL INFUSOVET CONNECT INFUSION PUMP
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Unlock: press any keypad, or click the screen, a reminder of unlock will be popped out, click .
Note!
The equipment will automatically unlock if there is high Level alarm.
7.2.5 Brightness
Click Brightnessto enter into display brightness setting interface. The brightness has 10 levels.
7.2.6 Night Mode
Click Night mode to enter into night mode switch setting interface to set the start and end time of the night mode and
the night brightness, at night, the system automatically adjusts the brightness to the User defi ned value.
7.2.7 Nurse Call
ClickNurse callto select function ON and OFF.
Note!
The nurse call function must be used with special cable.
The user shall not only depend to relay on the nurse call function as the main alarm notice mode, and shall
identify according to the equipment alarm and the animal state.
7.2.8 Nurse Call Alarm Level
Clicknurse call alarm levelto select different alarm levels.
7.2.9 Battery Capacity Display
Display of battery capacity under h:m or percentage status can be switched and title bar display changes accordingly.
7.3 Patient
ClickPatientin the main interface to enter into setting interface.
7.3.1 Patient Information
ClickPatientto enter into the patient information setting interface and set bed number, MRN, name, gender, age, body
weight, height.
7.4 Records
ClickRecordsin the main interface to enter into setting interface.
7.4.1 History Entries
ClickRecordsin the main interface to enter submenu, click theHistory entriesmenu item into history records query
interface. The equipment supports to save over 5,000 history records, and can display the event name, event date and time.
When it is full, the new records will cover the old records by turn.
The history record contains alarm information, treatment records and exhaust, cumulating clear, switch, standby operation
information.
7.4.2 Last Therapies
ClickRecordsin the main interface to enter submenu, click the “Last therapies” menu item into medical records query
interface.
1. This interface displays the latest 20 medical records, user may directly select it as the current infusion plan, after confi rming
the parameters, then start infusion.
2. The system can save 20 medical records at most, when it is full, the new records will cover the old records by turn.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer INFUSOVET CONNECT and is the answer not in the manual?

Eickemeyer INFUSOVET CONNECT Specifications

General IconGeneral
BrandEickemeyer
ModelINFUSOVET CONNECT
CategoryMedical Equipment
LanguageEnglish

Related product manuals