USER MANUAL INFUSOVET CONNECT INFUSION PUMP
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
9.7 Recycling
The normal service life of this equipment is 10 years, and depends on the use frequency and maintenance. The equipment
must be rejected after reaching the service life, please contact EICKEMEYER® or the dealer to get more detailed information.
1. The obsolete equipment may be returned to the original dealer or EICKEMEYER®.
2. The used lithium-ion polymer battery has the same treatment method, or according to the applicable laws and regulations.
3. Please handle according to the equipment rejecting fl ow of your medical mechanism.
10. BATTERY
This equipment is equipped with rechargeable lithium-ion polymer battery to ensure the normal infusion when the equipment
is moved or the external power supply is cut off.
When connecting external power supply, no matter the equipment is power on or not, the battery is charged. When charging,
the equipment screen displays the battery charging indication icon
When connecting external power supply, no matter the equipment is power on or not, the battery is charged. When charging,
. In case only built-in battery is adopted for
supplying power, and when the remained battery is less than 20%, please connect the equipment with external power supply
to charge the battery.
Warning!
Only the battery designated by EICKEMEYER® shall be adopted.
10.1 Check the Battery Performance
The performance of the built-in battery may drop according to the using duration, it is suggested to check the battery once a
month.
1. Disconnect the equipment from the animal, and stop all infusions.
2. Supply public power to the equipment to charge the battery for 10 h at least.
3. Supply power to the infusion pump only with battery, infusion at the rate of 25ml/h, test the time till the battery runs
down and the equipment is turned off.
4. If the battery power supply time is signifi cantly lower than the time stated in the specifi cation, consider replacing the
battery or contacting us.
10.2 Replaced the Battery
It is recommended to replace the battery every 2 years, it is suggested to replace the battery by the dealer or EICKEMEYER®.
Warning!
Untrained personnel are forbidden to replace the battery, otherwise it may cause the battery to burn, explode, leak
and cause personal injury.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer INFUSOVET CONNECT and is the answer not in the manual?

Eickemeyer INFUSOVET CONNECT Specifications

General IconGeneral
BrandEickemeyer
ModelINFUSOVET CONNECT
CategoryMedical Equipment
LanguageEnglish

Related product manuals