USER MANUAL INFUSOVET CONNECT INFUSION PUMP
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
9.2.1 Check the Appearance
1. The appearance of the equipment shall be clean and under good condition without crack and water leakage.
2. The buttons are fl exible and effective without invalid phenomenon; the sensitivity of the touch screen is normal.
3. The infusion pump door can be smoothly opened and closed, the safety clamp switch is under good condition.
4. The power wire is under good condition and installed tightly.
5. After connecting with external power supply, check whether the AC and DC indicators of the device and the battery
indicator are lit normally.
6. Adopt the accessories designated by EICKEMEYER®.
7. The environment meets the requirements.
9.2.2 Performance Check
Self-test and normal infusion function.
Alarm function normal.
Battery performance.
9.2.3 Maintenance Plan
The following check / maintenance items must be performed by the professional technician recognized by EICKEMEYER®. If
the following maintenances are necessary, please contact EICKEMEYER®. Please clean and disinfect the equipment before
testing or maintaining.
Maintenance Items Cycle
Safety check according to IEC 60601-1
Once every 2 years, please check after replacing the printed
circuit board assembly or the equipment is dropped or
knocked.
Preventive system maintenance items (pressure calibrate,
sensor calibrate, pump)
Once every 2 years, when the occlusion alarm, air bubble
alarm, or infusion accuracy is doubt to be abnormal
Brand of user-defi ned infusion apparatus, infusion accuracy
calibration
Using the equipment for the fi rst time, infusion apparatus
brand using for the fi rst time, reusing the equipment after
stopping for a very long period.
9.3 Add new Brand and Calibration
In the System submenu, click maintenance to enter into brand setting interface, create the consumables brand,
delete and calibrate the brand.
Warning!
It is suggested to contact EICKEMEYER® or local dealer, and customize and calibrate it by professional technician,
otherwise, it can’t guarantee the infusion accuracy.
Note!
The built-in brand of the system shall not be deleted.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer INFUSOVET CONNECT and is the answer not in the manual?

Eickemeyer INFUSOVET CONNECT Specifications

General IconGeneral
BrandEickemeyer
ModelINFUSOVET CONNECT
CategoryMedical Equipment
LanguageEnglish

Related product manuals