EasyManuals Logo

Eickemeyer INFUSOVET CONNECT User Manual

Eickemeyer INFUSOVET CONNECT
52 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #26 background imageLoading...
Page #26 background image
26
USER MANUAL INFUSOVET CONNECT INFUSION PUMP
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
6.2.5.3 Drip Mode
Under this mode, set the VTBI and drop rate, and the system will automatically calculate the infusion fl ow rate and time.
Note!
The ow rate under drip mode is calculated according to the specifi cation of the current infusion apparatus, before
adopting the drip mode, please confi rm that the specifi cation of the current infusion apparatus is accordant with
the specifi cation displayed in the interface title bar display, if it is not accordant, please contact the equipment
maintenance technician to modify, otherwise, it may cause serious deviation of fl ow rate.
6.2.5.4 Drug Library Mode
Nonemeans that the drug library mode is turned off. Click on the drug name and follow the instructions to enter the
infusion parameters.
DERS is suitable for this mode and the drug dose rate will be limited. If the cumulative dose exceeds the preset dose limit for
a certain period of time, the “drug dose overrun” alarm will be triggered.
Note!
This device supports customized drug information editing functions. Please contact the licensor if necessary.
6.2.5.5 Sequence Mode
Sequence mode means to infusion according to the set sequence after setting the rate and time of different sequence groups.
At most 5 sequence can be set in this mode.
6.2.6 Purge Air
To prevent air from entering the body, the air bubbles in the infusion set must be removed before infusion. Under the
parameters setting interface, Short pressBolusbutton to enter the exhaust interface, and exhaust according to the
interface instructions to clear the bubbles in the infusion line.
The purge total volume is not calculated in the Total Volume Infused.
Caution!
Before purge air, pls. confi rm the infusion line is not connected with the animal.
6.2.7 Set the Infusion Parameters
In each infusion mode, the user sets the infusion parameters through the touch screen. For the setting range of the infusion
parameters, see 2.5 Product Specifi cation.
6.2.8 Start Infusion
Connect IV tube with animal, confi rm infusion parameters, PressStartbutton, start infusion.
6.2.9 Changing Infusing Parameters during Infusion
During the infusion process, click the fl ow rate value on the running interface to reset the ow rate. After confi rmation, click
to continue the infusion.
Note!
Sequence mode does not support changing the fl ow rate during infusion.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer INFUSOVET CONNECT and is the answer not in the manual?

Eickemeyer INFUSOVET CONNECT Specifications

General IconGeneral
BrandEickemeyer
ModelINFUSOVET CONNECT
CategoryMedical Equipment
LanguageEnglish

Related product manuals