USER MANUAL INFUSOVET CONNECT INFUSION PUMP
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
• Only those infusion apparatus, line, infusion needle and other medical components that meet the local laws
and regulations and the requirements covered in and this User Manual can be adopted, it is suggested to adopt
the infusion apparatus with same brand as this equipment. It can’t ensure the infusion accuracy if the unsuitable
infusion line is adopted.
• It is not allowed to disassemble or refi t this equipment or use it for other purposes except normal infusion.
• No one is allowed to repair this equipment except EICKEMEYER® or the authorized repair technician of
EICKEMEYER®.
• Maintenance or replacement of spare parts is prohibited during the clinical use of the equipment.
• To avoid risk of electric shock, this equipment must only be connected to AC with Ground Protection.
1.2 Cautions
Caution!
• Before its fi rst use after purchase, or this equipment is not used for a long period, please charge the equipment
with AC power supply. If it is not fully charged, under power failure, the equipment can’t continue working with
built-in battery power supply.
• This equipment can not be used in the places with radiological installation or magnetic resonance equipment
as well as the places with high pressure oxygen therapy.
• Not to position ME Equipment to make it diffi cult to operate the disconnection device.
• The DC power supply is only suitable for applications where a backup power supply is required. Only use the
DC power supply line provided by EICKEMEYER®.
• Other devices near this equipment must meet corresponding EMC requirements, otherwise, it may infl uence the
performance of this equipment.
• Under general conditions, please use AC power supply as much as possible since it can prolong the service life
of the battery at a certain degree. When using AC power supply, ensure that the grounding wire is reliably
connected with the ground, and only the AC power wire attached with this equipment shall be adopted. The
built-in battery can only be used as the assistant power supply when the AC power supply can’t reliably
connected with the ground and is not under normal conditions (power failure or moving infusion).
• Before connecting this equipment with power supply, please keep the power socket and plug dry, and the
power voltage and frequency meet the requirements listed in the equipment label or this User Manual.
• The equipment is equipped with the audible and visual alarm system, and the red and yellow alarm indicators
will light on by turn to check if the alarm system can work normally, and the speaker makes the “beep” sound.
• Please keep the equipment away from the AC power socket for a certain distance to avoid fl uid / drug splashing
or dropping in the socket, otherwise, it may possibly cause short circuit.
• Please use the fl uid / drug after it has reached or nearly reached room temperature. When the fl uid / drug is
used at low temperature, the air which is dissolved in the fl uid / drug may cause more air bubbles and result in
frequent air bubble alarm.
• It is not allowed to press and operate the button with sharp object (such as pencil tip and nail), otherwise, it
may possibly cause early damage to button or surface fi lm.
• Please do not use the infusion tube for 8h at the same pumping position. Infusion tube may distort after using
for a long time and cause fl ow rate error. It is suggested to replace the pumping position or directly replace the
infusion tube every 8h.
• Please tightly close the fl ow rate adjuster of the infusion apparatus before taking out the infusion apparatus
to avoid liquid leakage.
• Under the condition of low fl ow rate infusion, please pay special attention on occlusion. The lower the infusion
fl ow rate, the longer the time of detecting occlusion, and it in turn may possibly cause a long time infusion stop
during this period.