USER MANUAL INFUSOVET CONNECT INFUSION PUMP
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
4.2.2 Install the Drop Sensor
1. Insert the drop sensor plug into the drop sensor port of this equipment and ensure tight connection.
2. Drop start should be above the line .
3. Liquid level should be below the line
Warning!
The fl uid / drug volume in the murphy’s dropper must be less than 1 / 3 of its volume.
The drop sensor shall be vertical.