USER MANUAL INFUSOVET CONNECT INFUSION PUMP
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
Warning!
When adopting uid / drug of high viscosity and the occlusion pressure is set at low level, it is possible that the
system will report occlusion alarm even when the line is not obstructed, under this condition, please carefully
observe the pressure indication icon in the display screen and infusion line, and rise the occlusion pressure if
needed.
When the occlusion pressure is set at high level, it may possibly cause the animal uncomfortable, after rising
the occlusion pressure, please carefully observe the condition of the animal, and immediately take measure if
there’s any abnormality.
Under the equipment fault state, the max pressure generated by the infusion line is 1,500mmHg. Under single
fault state, the max infusion volume is 2ml.
If not used for intravenous infusion, for example Intra-arterial infusion, TPN (Total Parenteral Nutrition) or EN
(Enteral Nutrition) treatment, occlusion level should be adjusted to higher levels.
Note!
The lowest pressure (50mmHg) limits the fl ow rate to ≤100ml/h, and the remaining 2 5 levels have no fl ow rate
limit.
When the line occlusion activates occlusion alarm, the system will automatically trigger anti-bolus function to drop the line
pressure and avoid additional impact bolus to the animal after contacting the occlusion. Liquid leakage will be less than
0.2ml, line pressure will be less than 300mmHg.
7.1.5 Bubble Detecting Level
ClickBubbles sizeto enter into air bubble size setting interface, move the long box to the preset level, confi rm and then
click .
The bubble sensitivity is 20µl.
Single bubble detection: A single bubble alarm is triggered when the individual bubble volume of the infusion tube reaches
the preset bubble detection alarm threshold. The individual bubble detection levels are detailed in the table below:
Standard software
Air Bubble detector level Alarm Threshold Value
Level 1 50µl
Level 2 100µl
Level 3 200µl
Level 4 400µl
Level 5 800µl

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer INFUSOVET CONNECT and is the answer not in the manual?

Eickemeyer INFUSOVET CONNECT Specifications

General IconGeneral
BrandEickemeyer
ModelINFUSOVET CONNECT
CategoryMedical Equipment
LanguageEnglish

Related product manuals