USER MANUAL INFUSOVET CONNECT INFUSION PUMP
T +49 7461 96 580 0 | F +49 7461 96 580 90 | export@eickemeyer.com | www.eickemeyer.com
9.4 Repair
Warning!
The maintenance of the equipment and the replacement of the components shall be carried out by professionals
recognized by EICKEMEYER® Special attention shall be paid to the detection of the power supply when the power
module is replaced. Observe whether there is a false alarm, connect the AC power supply, and the battery is charged
normally.
9.4.1 Normal Repair Process
Please contact EICKEMEYER® or authorized service personnel to repair if theres any fault, do not disassemble and repair the
equipment. After repair, please perform overall test for the equipment. EICKEMEYER® may provide the circuit diagram and
components list to the authorized repair technician if needed.
9.4.2 Maintenance for Long Term Store
If the equipment won’t be used for a long period, please pack the equipment in the package, and store it in the shade, cool
and dry place without direct sunlight.
The following operations are necessary for using it again:
1. Verify the fl ow rate accuracy to avoid unconformity between the infusion apparatus parameters in the equipment and the
actual parameters after it hasn’t be used for a long period or caused by other reasons, otherwise, it may cause infusion
error, infl uence the therapeutic effects and even cause medical negligence.
2. Perform air bubble and occlusion alarm test.
3. Test the battery discharging and charging duration to confi rm that the battery is also usable.
9.5 Equipment Components / Accessories
Warning!
Only the components and accessories designated by EICKEMEYER® shall be adopted, otherwise, it may possibly
damage the equipment or drop the equipment performance.
During the normal service life of the equipment, the battery and waterproof membranes are consumables, it is suggested to
replace them once every 2 years, please contact the dealer or EICKEMEYER® to replace them.
Standard Accessories
Two battery
Waterproof fi lm
Locking mechanism
Power cable
Optional Accessories
WiFi module
Drop sensor
DC Power cable
DB15 serial port communication wire
9.6 Production Date
Please refer to the label of the product.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Eickemeyer INFUSOVET CONNECT and is the answer not in the manual?

Eickemeyer INFUSOVET CONNECT Specifications

General IconGeneral
BrandEickemeyer
ModelINFUSOVET CONNECT
CategoryMedical Equipment
LanguageEnglish

Related product manuals