9. MAINTENANCE
9.1 Cleaning, Disinfecting and Sterilizing
Warning!
• Please cut off power supply and unplug the AC / DC power wire before cleaning the equipment.
• During cleaning and disinfecting, please keep the equipment horizontal and upwards to protect the equipment
and accessories from fl uid.
9.1.1 Cleaning
1. The daily maintenance is mainly to clean the housing and pump body. It is inevitable that fl uid / drug may fl ow in the
equipment during infusion. Some fl uid drug may corrode the pump and cause working fault. After infusion, please timely
clean the equipment, wipe it with moist and clean soft fabric, and then naturally dry it.
2. When cleaning the equipment interface, please wipe it with dry and soft fabric, confi rm the interface is dry before using.
3. Please do not soak the equipment in water. Although this equipment has certain waterproof function, when fl uid splashes
on the equipment, please check if it works normally, perform insulation and electric leakage test if needed.
9.1.2 Disinfecting
1. Disinfecting may possibly cause harm of certain degree to the equipment, it is suggested to disinfect the equipment if it
is needed.
Please disinfect the equipment with common disinfecting agent such as 70% ethanol, 70% isopropyl alcohol and so on.
Please follow the instructions of the disinfecting agent.
2. After disinfecting, wet the soft fabric with warm water, dry the fabric and then wipe the equipment with it.
3. Do not sterilize the equipment with high pressure steam sterilizer, do not dry the equipment with dryer or similar product.
Warning!
Please do not adopt Cidex OPA orthophthalaldehyde, methyl ethyl ketone or similar solvent, otherwise, it may
corrode the equipment.
9.2 Periodical Maintenance
Note!
• The medical mechanism shall set up complete maintenance plan, otherwise, it may possibly cause the
equipment malfunction or fault, and may possibly hurt the physical safety.
• In order to ensure the safe use and prolong the service life of the equipment, it is suggested to periodically
maintain and check it once every 6 months. Some items shall be maintained by the user, and some items shall
be maintained by the dealer of the equipment.
• Please timely contact EICKEMEYER® if the equipment is found defective.