Guidance and manufacture’s declaration – electromagnetic immunity
The Infusion pump is intended for use in the electromagnetic environment specifi ed below. The customer or the user of
the Infusion pump should assure that it is used in such an environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment
– guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 8 kV contact
± 15 kV air
Floors should be wood,
concrete or ceramic tile. If
fl oor are covered with
synthetic material, the
relative humidity should be
at least 30%.
Electrical fast transient / burst
IEC 61000-4-4
± 2 kV for power supply lines
+ 1 kV for input / output lines
± 2 kV for power supply lines
+ 1 kV for input / output lines
Mains power quality should
be that of a typical
commercial or hospital
environment.
Surge IEC 61000-4-5
± 1 kV line(s) to line(s)
± 2 KV line(s) to earth
± 1 kV line(s) to line(s)
± 2 kV line(s) to earth
Mains power quality should
be that of a typical
commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
40% U
T
(60% dip in U
T
)
for 5 cycles
70% U
T
(30% dip in U
T
)
for 25 cycles
Mains power quality should
be that of a typical
commercial or hospital
environment. If the user of
the Infusion pump requires
continued operation during
power mains interruptions, it
is recommended that the
Infusion pump be powered
from an uninterruptible
power supply or a battery.
Power frequency (50 / 60Hz)
magnetic fi eld IEC 61000-4-8
3 A/m 400A/m
Power frequency magnetic
fi elds should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment.
NOTE U
T
is the a.c. mains voltage prior to application of the test level.