E-ESD3000-E-Manual.doc 15/66
2 Safety
The ESD3000belongs to Safety class 1
2.1 Safety standard
The ESD3000 fulfils the requirements of the safety standards IEC 61010 for laboratory measurements
equipment „Safety requirements for electrical measuring, control and laboratory equipment“. Based on EN
61010 (IEC61010) the declaration of conformity to low voltage directive (LVD 73/23/EEC O.J.N° L77, 1973-
03-26) is given.
This manual is a integral part of the ESD3000 tester. The instructions
contained in the manual regarding operation and the test set up are to
be strictly observed
2.2 Climatic Conditions
The ESD3000 contains high voltage circuits in integrated form. EMC PARTNER only guarantees a correct
functioning of the tester ESD3000 and the associated accessories, if the ESD3000 is operated in the
climatic condition specified.
Temperature 15 °C to 35 °C Charging voltage up to 25 kV
Relative humidity 10 % to 60 % Charging voltage up to 25 kV
Atmospheric pressure 86 kPa to 106 kPa
(860 to 1060 mbar)
Charging voltage up to 25 kV
Atmospheric condition Voltages > 25 kV
20°C, RH 30%, pressure 101.3 kPa
Contact discharge not influenced by: direct solar radiation, rain or condense water, dust or larger
electro magnetic fields as specified in the EMC compatibility
chapter.
The ESD3000 should be operated in a dry, clean room. If for any reason water condenses in the ESD3000,
then no operation should be started before the tester is dry.
It is strictly forbidden to operate the ESD3000 in rooms with of gas explosion risk.
The high voltage of the ESD3000 generate sparks, which can ignite the gas.
People with heart pacemakers should not be in the vicinity of the test set up during operation.