Rosemount Model 8800C Vortex Flowmeter
D-6
E
Message Description
XMTR still busy. Flowmeter is running a computational or diagnostic routine and
cannot respond to the Model 268 instructions. Press the PREVIOUS
FUNCTION key to cancel.
XXXX=YYYY ERRIllegal value. Unacceptable value entered.
XXXX=YYYY ERR
Out of range.
The difference between LRV and URV is greater than the maximum
span allowed by the flowmeter. Enter a new value.
XXXX=YYYY ERRSpan too small. The difference between LRV and URV is less than the minimum span
allowed by the flowmeter. Enter a new value.
XXXX=YYYY ERRValue was too hi. See instructions for the selected parameter. Enter a new value that
is acceptable.
XXXX=YYYY ERRValue was too lo. See instructions for the selected parameter. Enter a new value that
is acceptable.
XXXX=YYYY ERRLRV too hi. LRV was set to a value above the URL. Enter a new value.
XXXX=YYYY ERRLRV too lo LRV was set to a value above the LRL Enter a new value.
XXXX=YYYY ERR
LRV and URV out limits.
Both the LRV and URV were outside the sensors range limits. Enter
a new value.
XXXX=YYYY ERRURV too hi. URV was set to a value above the URL. Enter a new value.
XXXX=YYYY ERRURV too lo. URV was set to a value below the LRL. Enter a new value.
XXXX=YYYY ERR268 data err. Flowmeter cannot interpret a Model 268 inquiry. Make note of
conditions and keystrokes leading to failure, and contact Rosemount
Service Center.
268 FailureNo. 1.
268 FailureNo. 2. Model 268 has detected bug in its software. Make note of conditions
and keystrokes leading to failure, and contact Rosemount Service
Center.
268 Test: FAIL. Model 268 is not functioning properly. Turn OFF and contact
Rosemount Service Center.