EasyManua.ls Logo

Endress+Hauser Cerabar M - Page 153

Endress+Hauser Cerabar M
234 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Cerabar M, Deltabar M, Deltapilot M Commissioning with the FF configuration program
Endress+Hauser 153
Calibration without reference pressure – dry calibration
Example:
In this example, the volume in a tank should be measured in liters. The maximum volume of
1000 liters corresponds to a level of 4.5 m. The minimum volume of 0 liters corresponds to
a level of 0.5 m since the device is mounted below the level lower-range value. The density
of the medium is 1 g/cm
3
.
Prerequisite:
The measured variable is in direct proportion to the pressure.
This is a theoretical calibration i.e. the height and volume values for the lower and upper
calibration point must be known.
CAUTION
!
Note Dependencies when setting parameters!
The values entered for Empty calibration/LOW_LEVEL_EASY and Full calib./
HIGH_LEVEL_EASY must be at least 1 % apart for the "Level easy pressure" level mode.
The value will be rejected, and a warning message displayed, if the values are too close
together. Further limit values are not checked; i.e. the values entered must be
appropriate for the sensor and the measuring task so that the measuring device can
measure correctly.
Due to the orientation of the device, there may be a shift in the measured value, i.e. when
the container is empty, the Primary Value/PRIMARY_VALUE parameter does not display
zero.
For information on how to perform position adjustment, see also ä 144,
"Pos. zero adjust/PRESSURE_1_ACCEPT_ZERO_INSTALL".
Description
1 Deltabar M: Before configuring the device for your
application, the pressure piping must be cleaned and
filled with medium.
A0031027
Fig. 46: Calibration without reference pressure – dry
calibration
A See Table, step 8.
B See Table, steps 10 and 11.
C See Table, steps 12 and 13.
2 Open the Pressure Transducer Block and set the
block mode to OOS.
3 Select the measuring mode if necessary:
Select the "Level height" option via the Primary Value
Type/PRIMARY_VALUE_TYPE parameter. Or:
4 Select the "Level" measuring mode via the Measuring
mode/OPERATING_MODE parameter.
Select the "In height" option via the Level selection/
LEVEL_ADJUSTMENT parameter.
5 By means of the Unit before Lin./OUT_UNIT_EASY
parameter, select a volume unit, here "l" for example.
6 Select a height unit by means of the Height unit/
HEIGHT_UNIT_EASY parameter, here "m" for
example.
7 Select the "Dry" option via the Calibration mode/
LEVEL_ADJUST_MODE_EASY parameter
8 By means of the Adjust density/
LEVEL_ADJUST_DENSITY_EASY parameter, enter a
density, here "1" "g/cm
3
" for example.
9 By means of the Empty calibration/
LOW_LEVEL_EASY parameter, enter a volume, here
0 l for example.
10 By means of the Empty height/
LEVEL_OFFSET_EASY parameter, enter a height,
here 0.5 m for example.
11 By means of the Full calib./HIGH_LEVEL_EASY
parameter, enter a volume, here 1000 l for example.
C
1000 l
4.5 m
B
0 l
0.5 m
r=1
g
cm
3
A

Table of Contents

Related product manuals