Operating menu and parameter description iTEMP TMT82
104 Endress+Hauser
HART message
Navigation Expert → Communication → HART info → HART message
Description Use this function to define a HART message which is sent via the HART protocol when
requested by the master.
User entry Up to 32 alphanumeric characters (letters, numbers and special characters)
Factory setting The device name
Hardware revision
Navigation Expert → Diagnostics → Device information → Hardware revision
Expert → Communication → HART info → Hardware revision
Description Use this function to display the hardware revision of the device.
SWRev
Navigation Expert → Communication → HART info → SWRev
Description Use this function to display the software revision of the device.
HART date code
Navigation Expert → Communication → HART info → HART date code
Description Use this function to define date information for individual use.
User entry Date in the format year-month-day (YYYY-MM-DD)
Factory setting 2010-01-01
"HART output" submenu
Assign current output (PV)
Navigation Expert → Communication → HART output → Assign current output (PV)