iTEMP TMT82 Operating menu and parameter description
Endress+Hauser 67
Description Use this function to select the sensor type for the sensor input in question.
Sensor type 1: settings for sensor input 1
Sensor type 2: settings for sensor input 2
Please observe the terminal assignment when connecting→  8,  17 the
individual sensors. In the case of 2-channel operation, the possible connection
options must also be observed.
Options A list of all the possible sensor types is provided in the "Technical data" section.→  44
Factory setting Sensor type 1: Pt100 IEC751
Sensor type 2: No sensor
Connection type n
Navigation Setup → Connection type n
Expert → Sensor → Sensor n → Connection type n
Prerequisite An RTD sensor must be specified as the sensor type.
Description Use this function to select the connection type for the sensor.
Options Sensor 1 (connection type 1): 2-wire, 3-wire, 4-wire
Sensor 2 (connection type 2): 2-wire, 3-wire
Factory setting Sensor 1 (connection type 1): 4-wire
Sensor 2 (connection type 2): 2-wire
2-wire compensation n
Navigation Setup → 2-wire compensation n
Expert → Sensor → Sensor n → 2-wire compensation n
Prerequisite An RTD sensor with a 2-wire connection type must be specified as the sensor type.
Description Use this function to specify the resistance value for two-wire compensation in RTDs.
User entry 0 to 30 Ohm
Factory setting 0
Reference junction n
Navigation Setup → Reference junction n
Expert → Sensor → Sensor n → Reference junction n