iTEMP TMT82 Operating menu and parameter description
Endress+Hauser 99
Call./v. Dusen coeff. A, B and C
Navigation Expert → Sensor → Sensor n → Linearization → Call./v. Dusen coeff. A, B, C
Prerequisite The RTD platinum (Callendar/Van Duse) option is enabled in the Sensor type parameter.
Description Use this function to set the coefficients for sensor linearization based on the
Callendar/Van Dusen method.
Factory setting • A: 3.910000e-003
• B: -5.780000e-007
• C: -4.180000e-012
Polynomial coeff. R0
Navigation Expert → Sensor → Sensor n → Linearization → Polynomial coeff. R0
Prerequisite The RTD poly nickel or RTD copper polynomial option is enabled in the Sensor type
parameter.
Description Use this function to set the R0 Value only for linearization of nickel/copper sensors.
User entry 40.000 to 1 050.000 Ohm
Factory setting 100.00 Ohm
Polynomial coeff. A, B
Navigation Expert → Sensor → Sensor n → Linearization → Polynomial coeff. A, B
Prerequisite The RTD poly nickel or RTD copper polynomial option is enabled in the Sensor type
parameter.
Description Use this function to set the coefficients for sensor linearization of copper/nickel resistance
thermometers.
Factory setting Polynomial coeff. A = 5.49630e-003
Polynomial coeff. B = 6.75560e-006
"Diagnostic settings" submenu
Calibration counter start