Levelflex FMP51, FMP52, FMP54 FOUNDATION Fieldbus Operating menu
Endress+Hauser 167
Factory setting • For Tank level (→ 166) = Partially filled: 0 mm (0 in)
• For Tank level (→ 166) = Fully flooded: 250 mm (9.8 in)
Additional information
A0013174
Dependence on the "Tank level" parameter
• Tank level (→ 166) = Partially filled:
In this case the Distance to upper connection parameter does not influence the
measurement. Thus, the default setting needs not to be changed.
• Tank level (→ 166) = Fully flooded:
In this case enter the distance D
U
between the reference point and the lower edge of the
upper connection.
DC value
Navigation Setup → DC value
Prerequisite The device has the "interface measurement" application package
9)
.
Description Specify relative dielectric constant ε
r
of the upper medium (DC
1
).
User entry 1.0 to 100
Factory setting 2.0
9) Product structure: Feature 540 "Application Package", Option EB "Interface measurement"