Liquistation CSF48 Communication
105
7.2 Service interface
You can connect the device to a computer via the service interface and configure it using
"Fieldcare". Furthermore, configurations can also be saved, transferred and documented.
7.2.1 Connection
Connect the service connector to the interface on the Liquiline basic module and connect it
to the Commubox.
Via the USB port, connect the Commubox to the computer running the Fieldcare software.
a0014953
Fig. 15: Connection overview
7.2.2 Creating the data connection
Start Fieldcare.
Establish the connection to the Commubox.
For this purpose, under "Host PC" click "Add device" and select "CDI Communication FXA291".
Click "Configuration" to select the serial interface for FXA291 and to set the baud rate
(115200).
Click "Link connection" to start communicating with the FXA291 ( are green).
Click "Add device" again and select your device type.
Click "Link connection" to start communicating with the device ( are green).
Select "Online parameterize" to be able to communicate with the device online.
You can now start online configuration via the DTM.
Online configuration competes with onsite operation, i.e. each of the two options blocks the
other one. Either side is able to prevent access from the other side.