Liquistation CSF48 Inputs: pH/ORP Incl. combi sensors
115
10 Inputs: pH/ORP Incl. combi sensors
10.1 Basic settings
10.1.1 Sensor identification
10.1.2 Main value
10.1.3 Damping
--> "Inputs: General/Frequently occurring functions" section ä 108
10.1.4 Manual hold
--> "Inputs: General/Frequently occurring functions" section ä 108
Path: Menu/Setup/Inputs/<Sensor type>
Function Options Info
Channel Options
•On
•Off
Factory setting
On
On
The channel display is switched on in the measuring
mode
Off
The channel is not displayed in the measuring mode,
regardless of whether a sensor is connected or not.
Sensor type Read only
(Only available if a sensor is
connected)
Connected sensor type
Order code Order code of the connected sensor
Path: Menu/Setup/Inputs/pH or Redox or pH/Redox
Function Options Info
Main value Options
• pH(only pH sensor)
•mV
• % (only ORP sensor)
•Redox mV (onlycombi
sensor)
• Redox % (only combi sen-
sor)
• rH (only combi sensor)
Factory setting
pH (pH sensor and combi
sensors)
mV(ORP sensor)
Select how the main measured value should be dis-
played.
You can display the main measured value of a pH sensor
as a pH value or as a raw value in mV.
If using an ORP sensor, here you decide which ORP mode
to use: mV or %. If you have connected a combi sensor,
you can also select the rH value.
Subsequent configuration options depend on the option
selected here.