Inputs: Conductivity Liquistation CSF48
124
11 Inputs: Conductivity
11.1 Basic settings
11.1.1 Sensor identification
11.1.2 Damping
--> "Inputs: General/Frequently occurring functions" section ä 108
11.1.3 Manual hold
--> "Inputs: General/Frequently occurring functions" section ä 108
11.1.4 Operating mode and cell constant
Path: Menu/Setup/Inputs/<Sensor type>
Function Options Info
Channel Options
•On
•Off
Factory setting
On
On
The channel display is switched on in the measuring
mode
Off
The channel is not displayed in the measuring mode,
regardless of whether a sensor is connected or not.
Sensor type Read only
(Only available if a sensor is
connected)
Connected sensor type
Order code Order code of the connected sensor
Path: Menu/Setup/Inputs/Conductivity
Function Options Info
Operating mode Options
Conductivity
•Resistance
(only Cond c)
Concentration
(only Cond i)
•TDS
(only Cond c)
Factory setting
Conductivity
Alternatively to the conductivity, you can also measure
the resistivity and the total dissolved solids (TDS)
parameter with a conductive conductivity sensor.
Alternatively to the conductivity, you can determine the
concentration of the medium with an inductive
conductivity sensor.
TDS
TDS stands for all the organic and inorganic substances
in the water in ionic, molecular or microgranular (<2
μm) form.
Compared with laboratory methods (gravimetric analy-
sis), TDS measurement via the conductivity value deli-
vers a maximum measured error of less than 10%.
Cell constant Read only
(Only available if a sensor is
connected)
The cell constant of the connected sensor is displayed
(--> sensor certificate)