Inputs: Conductivity Liquistation CSF48
130
11.2 Extended setup
11.2.1 Temperature format
11.2.2 Cleaning
--> "Inputs: General/Frequently occurring functions" section ä 108
11.2.3 Diagnostics settings
This menu branch is used for specifying warning limits, and for defining whether and how dia-
gnostics tools should be used.
The associated diagnostics code is displayed for every setting.
Process check system
--> "Inputs: General/Frequently occurring functions" section ä 108
Path: Menu/Setup/Inputs/Conductivity
Function Options Info
Temp. comp. mode
(Compensation=one of the
user tables)
Options
• Conductivity
• Coeff. Alpha
Factory setting
Conductivity
Conductivity
You specify the temperature, conductivity and
uncompensated conductivity. Recommended for large
measuring ranges and small measured values.
Coeff. Alpha
As the value pairs, you specify an alpha value and the
related temperature.
Table name
(Compensation=one of the
user tables)
Customized text, 16
characters
Assign a meaningful name to the selected table.
Edit table
(Compensation=one of the
user tables)
• Temperature
• Conductivity
• Temperature comp. cond.
or
• Temperature
• Coefficient alpha
Maximum number of rows: 25
The type of table depends on the option under "Temp.
comp. mode".
Path: Menu/Setup/Inputs/<Sensor type>/Extended setup
Function Options Info
Temperature format Options
•#.#
•#.##
Factory setting
#.#
Select how many decimal places should be used to
display the temperature.