Inputs Liquistation CSF48
24
3Inputs
Liquistation CSF48 has 2 binary inputs and 2 current inputs as standard. All inputs are galvani-
cally isolated from one another.
3.1 Binary inputs
The binary inputs are used to control the sampler using external signals.
With the CSF48, the auxiliary voltage of 24 V DC from the terminal block in the connection
compartment of the sampler can be used for floating contacts (see BA00443C "Commissioning").
Path: Menu/Setup/Inputs
Function Options Info
Binary input S:x
Mode Options
•Off
•On
Factory setting
Off
Switches the function on or off
Input mode Options
•Flow rate
•Rainfall
• External event
•Armature end position
detection (only for version
with sampling assembly)
Factory setting
Flow rate
• Pulse input for connected flowmeters or rain gages
• Control of sampling functions via external signals
If Input mode Flow rate is selected:
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Preselect the level change of the signal.
Unit Options
•m³
•l
•cf
•gal
Factory setting
m³
Select the unit.
Meas. value format Factory setting
#.#
Specify the number of decimal places for the flow.
1 Impulse = 0 to 1000 m
3
Factory setting
10 m
3
Definition of the pulse value, limits are calculated
depending on the unit