Inputs: Turbidity and solids Liquistation CSF48
152
14 Inputs: Turbidity and solids
14.1 Basic settings
14.1.1 Sensor identification
14.1.2 Application
The sensor is precalibrated on leaving the factory. As such, it can be used in a wide range of
applications (e.g. clear water measurement) without the need for additional calibration. The
factory calibration is based on a three-point calibration of a reference sample.
The factory calibration cannot be deleted and can be retrieved at any time. All other calibrations
-
performed as customer calibrations - are referenced to this factory calibration.
Calibration data records are saved under an individual name. You can add your own data
records during each calibration. These are then available for selection under "Application".
14.1.3 Damping
--> "Inputs: General/Frequently occurring functions" section ä 108
Path: Menu/Setup/Inputs/<Sensor type>
Function Options Info
Channel Options
•On
•Off
Factory setting
On
On
The channel display is switched on in the measuring
mode
Off
The channel is not displayed in the measuring mode,
regardless of whether a sensor is connected or not.
Sensor type Read only
(Only available if a sensor is
connected)
Connected sensor type
Order code Order code of the connected sensor
Path: Menu/Setup/Inputs/Turbidity
Function Options Info
Application type Options
Clear water
Solid
Factory setting
Clear water
Preselection for saved calibration data records
Application Depends on the sensor Select a saved calibration data record