Liquistation CSF48 Inputs: Interface
177
18 Inputs: Interface
18.1 Basic settings
The CUS71DOUS71D is not detected automatically. It must be selected manually (Current
sensor). During initial commissioning, data are recorded and calculated for 3 to 5 minutes
before a measured value is displayed.
18.2 Manual hold
--> "Inputs: General/Frequently occurring functions" section ä 108
Path: Menu/Setup/Inputs/Ultrasonic interface
Function Options Info
Sensor operation Options
•Scan for memosens
sensor
• Current sensor
Factory setting
Current sensor
Scan for memosens sensor
Searches for Memosens sensors
Current sensor
Connected sensor is used
Wiper function Options
•Off
•On
Factory setting
On
Only for sensor version with wiper function
Wiper timing 1 to 240 min
Factory setting
10 min
Only for sensor version with wiper function
Turbidity measurement Options
•Off
•On
Factory setting
On
Only for sensor version with integrated turbidity
measurement and wiper function
Turbidity unit Options
•FNU
•NTU
Factory setting
FNU
Only for sensor version with integrated turbidity
measurement and wiper function