Liquistation CSF48 General settings
13
2.5 Configuring the sampling depending on the device version
The list of functions displayed depends on the device version selected.
Functions marked
1)
are available in device versions with a vacuum pump.
Functions marked
2)
are available in device versions with a peristaltic pump.
Functions marked
3)
are available in device versions with a distribution arm drive.
Functions marked
4)
are available in device versions with a sampling assembly.
Path: Menu/Setup/General settings
Function Options Info
Sampling
Number of bottles Choice of all possible bottle
combinations
The bottle configuration you ordered is preset in
the device.
Bottle volume 0 to 100000 ml
Factory setting
Depends on the bottle configu-
ration
Distribution reference
3)
Options
Pre sampling
Pre bottle change
Pre program start
Factory setting
Pre sampling
The distribution arm goes through a reference point
depending on the option selected.
Pre sampling:
This means that the distribution arm performs a
reference test before each individual sampling.
Pre bottle change:
This means that the distribution arm performs a
reference test in every subprogram.
Pre program start:
This means that a single reference test is performed
before the program starts.
Power failure Options
Resume program
•Stop program
Factory setting
Resume program
Decide how the sampler should react when it is
energized after a power failure.
Resume program:
Time and flow-paced
The program calculates the omitted samples and
enters them in the logbook as failed. When the
program is restarted, it continues where it was
interrupted.
•Flow-paced
No samples are entered in the logbook during
the power failure. When the program is restar-
ted, it continues where it was interrupted.
Sample retries
1), 2), 3)
0 to 3
Factory setting
0
If sampling is started and no sample is drawn in,
sampling can be repeated up to 3 times.
Sampling delay 0 to 99 s
Factory setting
0 s
The start of the sampling cycle can be delayed by up
to 99 s. The binary output is switched without any
delay.