General settings Liquistation CSF48
14
Liquid detection
1)
Options
•Automatic
Semi automatic
Factory setting
Automatic
If "Semiautomatic" is selected, the purge times and
intake times can be defined separately.
Dosing volume
1), 4) 1)
20 to 350 ml
Factory setting
200 ml
4)
10 to 1000 ml
Factory setting
200 ml
1)
Adjust the dosing tube in the dosing chamber to
change the dosing volume. The level in the bottle is
calculated using the set dosing volume.
4)
If the sampling assembly CSA420 is used, 10, 30
or
50 ml are the only values that are permitted.
Conductive sensor
1)
Options
Low sensitivity
Medium sensitivity
High sensitivity
Factory setting
Medium sensitivity
The switching behavior can be set with the liquid
detection function. Use the high sensitivity setting
if the sample has a low conductivity, for example.
Dosing chamber
1)
Options
Dose without pressure (A)
Dose with pressure (B)
Factory setting
Dose without pressure (A)
Dosing with pressure e.g. in conditions with low
suction heights and slight counterpressure or low
volumes.
Liquid detection
2)
Options
•Automatic
Semi automatic
•Off
Factory setting
Automatic
If "Semiautomatic" is selected, the purge times and
intake times can be defined separately.
Off:
The definition of the purge times and intake times
is completely time-controlled.
Automatic:
The last intake time determined is the new purge
time.
Semi automatic:
If the suction heights tend to vary greatly.
Rinse cycles
2)
0 to 3
Factory setting
0
The suction line is rinsed with the sample up to 3
times.
Safety interlock
2)
(optional) Options
•Off
•On
Factory setting
Off
If the peristaltic pump is opened, the safety inter-
lock stops all the functions.
Samping time
4)
5 to 20 s
Factory setting
5 s
Enter the time during which the injected piston is
located in the sample.
Path: Menu/Setup/General settings
Function Options Info