Liquistation CSF48 Inputs: Chlorine
145
13.1.5 Unit
13.2 Extended setup
13.2.1 Measured value formats
13.2.2 Medium compensation (in the process)
Path: Menu/Setup/Inputs/Chlorine
Function Options Info
Unit
Main value="Concentration
liquid"
Options
•mg/l
μg/l
•ppm
•ppb
Factory setting
mg/l
Path: Menu/Setup/Inputs/DO or Chlorine/Extended setup
1)
1) In the case of chlorine, the sequence of the two menu functions is reversed
Function Options Info
Main value format Options
•#.#
•#.##
#.###
•#
Factory setting
#.##
Specify the number of decimal places for displaying the
main measured value.
Temperature format Options
•#.#
•#.##
Factory setting
#.#
Select how many decimal places should be used to
display the temperature.
Path: Menu/Setup/Inputs/Chlorine/Extended setup
Function Options Info
Medium comp. (pH) Options
•Off
•On
Factory setting
On
Off
The concentration measured value is calculated as HClO
(=free available chlorine).
On
The pH value is used to calculate a cumulative
concentration value from HClO and ClO
-
(=total
chlorine).