Inputs: Chlorine Liquistation CSF48
146
13.2.3 Cleaning
--> "Inputs: General/Frequently occurring functions" section ä 108
13.2.4 Calibration settings
Calibration timer and calibration expiration date
--> "Inputs: General/Frequently occurring functions" section ä 108
Stability criteria
You define the permitted measured value fluctuation which must not be exceeded in a certain
timeframe during calibration.
If the permitted difference is exceeded, calibration is not permitted and is aborted automatically.
Mode
Medium comp. (pH)="On"
Options
Fixed value
•Measured value
Factory setting
Fixed value
Decide whether you want to specify a fixed pH value for
calculating the total chlorine or whether the measured
value of a pH sensor attached to another input should be
used.
Fixed pH
Mode="Fixed value"
4.00 to 9.00 pH
Factory setting
7.20 pH
Useful for media with a constant pH value
Enter the pH value of your medium which you
determined with a reference measurement.
Associated pH-sensor
Mode="Measured value"
Select the pH sensor
Factory setting
None
Preferred method for media with varying pH values
Select the sensor input with the connected pH sensor.
The measured value of the sensor is then continuously
used to calculate the total chlorine.
Temp. compensation Options
•Off
•Automatic
•Manual
Factory setting
Automatic
Decide whether and how you want to compensate the
medium temperature:
•No compensation
Automatically using the temperature sensor of your
sensor
Manually by entering the medium temperature
Medium temperature
(Temp.
compensation=Manual)
-5.0 to 50.0 ˚C
(23.0 to 122.0 ˚F)
Factory setting
20.0 ˚C (68 ˚F)
Enter the temperature of your medium.
Path: Menu/Setup/Inputs/Chlorine/Extended setup
Function Options Info