Liquistation CSF48 General settings
9
2.4 Logbooks
Logbooks record the following events:
• Calibration/adjustment events
• Operation events
• Diagnostic events
•Programming events
Here you define how the logbooks should store the data.
In addition, you are also able to define individual data logbooks. Assign the logbook name and
select the measured value to be recorded. You can configure the recording rate (Scan time)
individually for every data logbook.
More information on the logbooks is provided in BA00463C "Maintenance & diagnostics", "Dia-
gnostics menu" section.
Path: Menu/Setup/General settings/Logbooks
Function Options Info
Logbook ident Customized text Part of the file name when exporting a logbook
Event logbook Options
•Off
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
All diagnostic messages are recorded
Ring buffer
If the memory is full, the latest entry automatically
overwrites the oldest entry.
Fill up buffer
If the memory is 80% full, the device displays a diag-
nostic message.
If the memory is full, there is an overflow, i.e. no new
values can be saved. The controller displays a
corresponding diagnostic message. The memory then
has to be cleared manually.
Logbook program Options
•Off
• Ring buffer
• Fill up buffer
Factory setting
Ring buffer
All program messages are recorded
Ring buffer
If the memory is full, the latest entry automatically
overwrites the oldest entry.
Fill up buffer
If the memory is 80% full, the device displays a diag-
nostic message.
If the memory is full, there is an overflow, i.e. no new
values can be saved. The controller displays a
corresponding diagnostic message. The memory then
has to be cleared manually.
Overflow warnings
Event logbook="Fill up buffer"
Calibration logbook Options
•Off
•On
Factory setting
Off
Decide whether you want to receive a diagnostic
message from the controller in the event of memory
overrun of the logbook in question.
Diagnostic logbook
Configuration logbook