Liquistation CSF48 Inputs: Oxygen
143
Diagnostic behavior
--> "Inputs: General/Frequently occurring functions" section ä 108
12.2.7 Tag control
--> "Inputs: General/Frequently occurring functions" section ä 108
12.2.8 Sensor replacement
--> "Inputs: General/Frequently occurring functions" section ä 108
12.2.9 Data processing factory setting
--> "Inputs: General/Frequently occurring functions" section ä 108
12.2.10 Sensor factory setting (only COS61D)
--> "Inputs: General/Frequently occurring functions" section ä 108
(Operation above second
specified nA value)
Only amperometric sensors, sensor-specific limit
– COS22D: >50 nA
– COS51D: >160 nA
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
185 "Operating time" (COS22D)
186 "Operating time" (COS51D)
Operation < 25 μs Only optical sensors (μS = fluorescence decay time, raw
value of optical measurement)
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
181 "Operating time"
Operation > 40 μs Only optical sensors
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
182 "Operating time"
Path: Menu/Setup/Inputs/DO/Extended setup/Diagnostics settings
Function Options Info