Inputs: SAC Liquistation CSF48
160
Path: Menu/Setup/Inputs/SAC/Extended setup/Diagnostics settings
Function Options Info
Limits operating hours Specify your limit values for monitoring the number of
operating hours under extreme conditions.
The range of adjustment for the operating hours alarm and warning limits is generally 1 to 50000 h.
Function Options
•On
•Off
Factory setting
Off
On
The operation of the sensor under extreme conditions is
monitored, recorded in the sensor and diagnostics
messages are displayed on the controller.
Off
No diagnostics messages. However, the time the sensor
operates under extreme conditions is recorded in the
sensor and can be read in the sensor information in the
diagnostics menu.
Operating time Total operating time of the sensor
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
199 "Operating time"
The names of the menu functions in brackets depend on the sensor specification. For this reason, they cannot be
specified here.
(Operation below specified temperature limit, e.g. < 5 ˚C)
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
935 "Process temp."
(Operation above specified temperature limit, e.g. > 50 ˚C)
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
934 "Process temp."
(Operation below specified limit value, e.g. < 50 mg/l)
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
170 "Process value"
(Operation above specified limit value, e.g. > 200 mg/l)
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
169 "Process value"
Filter change
Warning limit Factory setting
10000 h
Diagnostics code and associated message text:
157 "Filter change"
Alarm limit Factory setting
15000 h
Diagnostics code and associated message text:
161 "Filter change"