Programming Liquistation CSF48
38
Options:
- 1x - PE direct distribution
- 2x - PE direct distribution
- 4x - PE direct distribution
1)
- 4x - glass direct distribution
- 12x - PE direct distribution
- 12x - PE/glass distributor plate
- 24x - PE direct distribution
- 24x - PE/glass distributor plate
- 6x+1x - PE/glass distributor
plate
- 6x+2x - PE+PE distributor
plate
- 6x+2x - PE+glass distributor
plate
- 12x+1x - PE/glass distributor
plate
- 12x+2x - PE+PE distributor
plate
- 12x+2x - PE+glass distributor
plate
- 12x+6x PE direct distribution
- 12x+6x PE/glass distributor
plate
1)
Use this option for bottle configuration "W" (4 x
5000 ml Schott bottles, DURAN GLS 80).
Bottle volume 0 to 100000 ml
Factory setting
30000 ml
Set the bottle volume.
The preset value depends on the bottle configura-
tion configured. The bottle volume is always 30 l for
individual containers.
In the case of asymmetric distribution, e.g.
6x3l+2x13l, you can set the bottle
volume on the left and right in the menu
items that follow.
Sampling mode Options
•Time paced CTCV
•Flow paced VTCV
Time/flow paced CTVV
Factory setting
Time paced CTCV
The following functions depend on the option selec-
ted.
These versions are illustrated individually in the fol-
lowing section to provide a clearer understanding of
the options.
Time paced CTCV:
A constant sampling volume is taken at steady
intervals.
Flow paced VTCV:
A constant sampling volume is taken at variable
intervals.
Time/flow paced CTVV (only for version with
peristaltic pump):
A variable sampling volume is taken at steady inter-
vals.
Path: Menu/Setup/Sampling programs