Programming Liquistation CSF48
42
Time interval 00-00:02 to 31-00:00
DD-HH:MM
Factory setting
00-01:00 DD-HH:MM
Set the time (days, hours and minutes) after which
the system should change to the next bottle.
Multiple bottles 0 to 23
The configuration
options depend on the
current number of
bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to separate
bottles.
Bottle synchronization Options
•None
1. bottle change time
1. Time of change + bottle
number
Factory setting
None
None:
The time of sampling and bottle change are not
synchronized.
1. bottle change time:
Sampling starts with the first bottle.
Set the synchronization time.
1. Time of change + bottle number:
Each bottle is assigned to a specific fill time.
Set the synchronization time and the weekday.
If Bottle change mode External signal is selected:
Bottle change signal input Options
No bottle change input
configured
Binary input S:x
Factory setting
No bottle change input
configured
The bottle change input can be configured under
Inputs .
Multiple bottles 0 to 23
The configuration
options depend on the
current number of
bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to separate
bottles.
Start condition Options
•Immediate
•Date/time
Factory setting
Immediate
The sampling program can be started either
immediately or at a specific, configurable time.
If Start condition Date/time is selected:
Start date 01.01.2000 to 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program. The
format depends on the option configured under
general settings.
Path: Menu/Setup/Sampling programs/Setup program/New/Basic