Liquistation CSF48 Programming
55
Settings with the Basic program type with multiple bottles
Sampling mode = "External signal"
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Sampling volume 10 to 1000 ml
Factory setting
100 ml
Enter the sample volume.
Sampling signal input Options
No sampling input
configured
Factory setting
No sampling input configured
Select the input for the sampling signal.
The fieldbus must be configured for this function.
The sampling input can be configured under Inputs
.
Bottle change mode Options
•Number of samples
•Time
•External signal
Factory setting
Number of samples
The bottle can be changed either after a specific
number of samples, after a time or by an external
signal.
If Bottle change mode Number of samples is selected:
Samples per bottle 1 to 9999
Factory setting
1
Set the number of samples.
If Bottle change mode Time is selected:
Time interval 00-00:02 to 31-00:00
DD-HH:MM
Factory setting
00-01:00 DD-HH:MM
Set the time (in days, hours and minutes) after which
the system should change to the next bottle.
If Bottle change mode External signal is selected:
Bottle change signal input Options
No bottle change input
configured
Factory setting
No bottle change input
configured
The bottle change input can be configured under
Inputs .
Multiple bottles 0 to 23
The configuration
options depend on the
current number of
bottles
Factory setting
0
Multiple bottles:
"Simultaneous" transfer of two samples to separate
bottles.