Liquistation CSF48 Outputs
83
Event Options
Program enabled
End of program
Sampling start
End of sampling
•Dosing
Sampling cycle
Bottle change
•External stop
•No sample
Sub program enabled
Factory setting
Sampling cycle
Program enabled:
A permanent signal is switched when the sampling
program starts.
End of program:
A pulse or permanent signal is switched when the
sampling program ends.
Sampling start:
A pulse is switched when a sample is taken.
End of sampling:
A pulse is switched when sampling has ended.
Dosing:
A pulse is switched while the device doses a sample.
Sampling cycle:
The output signal is switched for the duration of the
sampling cycle.
Bottle change:
A pulse is switched when a bottle is changed.
External stop:
A pulse is switched when an external stop is
performed.
No sample:
The output signal is switched if no sample was taken.
Sub program enabled:
The output signal is switched if this subprogram is
active.
If Function: Limit value is selected:
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Select the level change of the signal
Source of data Options
•None
Limit switch 1-8
Factory setting
None
Select the limit switch via which the status of the relay
is to be output.
The limit switches are configured in the
"Setup/Additional functions/LimitSwitch" menu.
If Function: Diagnostics message is selected:
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Select the level change of the signal
Path: Menu/Setup/Outputs