Liquistation CSF48 Outputs
83
Event Options
• Program enabled
• End of program
• Sampling start
• End of sampling
•Dosing
• Sampling cycle
• Bottle change
•External stop
•No sample
• Sub program enabled
Factory setting
Sampling cycle
Program enabled:
A permanent signal is switched when the sampling
program starts.
End of program:
A pulse or permanent signal is switched when the
sampling program ends.
Sampling start:
A pulse is switched when a sample is taken.
End of sampling:
A pulse is switched when sampling has ended.
Dosing:
A pulse is switched while the device doses a sample.
Sampling cycle:
The output signal is switched for the duration of the
sampling cycle.
Bottle change:
A pulse is switched when a bottle is changed.
External stop:
A pulse is switched when an external stop is
performed.
No sample:
The output signal is switched if no sample was taken.
Sub program enabled:
The output signal is switched if this subprogram is
active.
If Function: Limit value is selected:
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Select the level change of the signal
Source of data Options
•None
• Limit switch 1-8
Factory setting
None
Select the limit switch via which the status of the relay
is to be output.
The limit switches are configured in the
"Setup/Additional functions/LimitSwitch" menu.
If Function: Diagnostics message is selected:
Signal slope Options
•Low-High
•High-Low
Factory setting
Low-High
Select the level change of the signal
Path: Menu/Setup/Outputs
Function Options Info