Liquistation CSF48 Outputs
89
5.3.3 Outputting the status of a cleaning function
Function="Diagnostics"
Function Options Info
Operating mode Options
• as assigned
•Namur M
•Namur S
•Namur C
•Namur F
Factory setting
as assigned
as assigned
If this option is selected, the diagnostics messages which
you have individually assigned to the relay are output via
the relay.
Namur M... F
If you decided to use one of the Namur classes, all the
messages that are assigned to the individual class are
output via the relay.
You can also change the Namur class assignment for
every diagnostics message.
(Menu/Setup/General settings/Diagnostics/Device
behavior or Menu/Setup/Inputs/../Diagnostics
settings/Diag. behavior)
Attributed diagnostic
messages
Operating mode="as assigned"
Read only
List of diagnostic messages
All the messages assigned to the relay output appear on
the display.
You do not have the option of editing the information
here.
Function="Cleaning"
Function Options Info
Assignment Options
•None
• Cleaning 1 - Water
• Cleaning 1 - Cleaner
• Cleaning 2 - Water
• Cleaning 2 - Cleaner
• Cleaning 3 - Water
• Cleaning 3 - Cleaner
• Cleaning 4 - Water
• Cleaning 4 - Cleaner
Factory setting
None
Here you can specify how a cleaning function should be
displayed for the alarm relay.
You can define the cleaning programs under:
Menu/Setup/Additional functions/Cleaning.