Liquistation CSF48 Operation
Endress+Hauser 107
Settings with a flow-paced Basic program
Settings with the Basic program type with 1 bottle
Sampling mode = "Flow paced VTCV"
Path: Menu/Setup/Sampling programs/Setup program/New/Basic
Function Options Info
Flowmeter input Options
• No flow input configured
• Binary input S:x
• Current input S:x
Factory setting
No flow input configured
Select the flow input.
The binary input or the current input must be
configured for this function.
Only the inputs configured as a flow input are
displayed.
Sampling interval (for
versions with vacuum pump or
peristaltic pump)
1.000 to 9999.000 m³
Factory setting
10,000 m³
Set the sampling interval. The unit and the number of
decimal places are displayed as configured under
Setup/Inputs.
Dosing volume (for version
with vacuum pump or
sampling assembly)
Sampling volume (for version
with peristaltic pump)
Vacuum pump:
20 to 350 ml
Peristaltic pump:
10 to 10000 ml
Sampling assembly:
10 to 1000 ml
Factory setting
Vacuum pump:
200 ml
Peristaltic pump:
100 ml
Sampling assembly:
200 ml
Set the dosing volume or the sampling volume.
The volume is taken from the setup in the version
with a vacuum pump or sampling assembly.
The dosing accuracy and the repeatability of a sample
volume < 20 ml can vary, depending on the specific
application.
Multiplier
(only for version with vacuum
pump or sampling assembly)
1 to 10
Factory setting
1
You can change the sampling volume using the
multiplier.
For example, if the dosing volume is set to 200 ml, set
the multiplier to 2 for a sampling volume of 400 ml. 2
samples are taken in succession when sampling.
Bottle change mode Options
•Number of samples
•Time
•External signal
Factory setting
Number of samples
The bottle can be changed after a specific number of
samples, after a time or by an external signal.
If Bottle change mode Number of samples is selected:
Samples per bottle 1 to 9999
Factory setting
1
Set the number of samples.