Programming example: Program switchover
The following example explains how to program a program change with flow-paced sampling
as the daily average sample from Monday to Friday and time-paced sampling to a composite
container on Saturday and Sunday.
Path: Menu/Setup/Sampling programs/Setup program/New
Function Entry for sample program
Standard
Program name TDVT 6+20
Bottle configuration 6x + 1x
Bottle volume left 3000 ml
Bottle volume right 20000 ml
Start condition Date/time
Start date DD.MM.YYYY e.g. 03.06.2010
Start time HH:MM:SS e.g. 12:00:00
Stop condition Continuous
Setup subprogram
Programpart MO - FR
"SAVE" - saves the "MO - FR" subprogram
Setup subprogram
Programpart SAT - SUN
"SAVE" - saves the "SAT - SUN" subprogram
"ESC" - the main program is displayed
Bottle assignment Statical bottle assignment
Setup subprogram: MO - FR
Edit
Sampling mode Flow paced VTCV
Flowmeter input Current input S:1
Sampling interval 1000 m³
Sampling volume 80 ml
Bottle change mode Time
Time interval 04:00
Multiple bottles 0
Enable subprogram Multiple date