Operation Liquistation CSF48
130 Endress+Hauser
Settings for the Advanced program
Path: Menu/Setup/Sampling programs
Function Options Info
Setup program
New A list of all the programs created is displayed. For this
reason, it is often helpful to add an "A" for Advanced
in the program name.
Advanced
Program name Customized text Use a distinct name for your sampling program. The
program name can be up to 16 characters long.
Bottle configuration Choice of all possible bottle
combinations
The ordered bottle configuration is preset or the
configuration selected in the setup is displayed.
Bottle volume 10 to 100000 ml
Factory setting
30000 ml
Set the bottle volume.
The preset value depends on the bottle configuration
configured. The bottle volume is always 30 l for
individual containers.
In the case of asymmetric distribution, e.g.
6x3l+2x13l, you can set the bottle volume on the
left and right in the menu items that follow.
Start condition Options
•Immediate
•Date/time
•Volume
•External start
•External duration
• Fieldbus (optional)
Factory setting
Immediate
Immediate
The sampling program is started immediately.
Date/time
The sampling program is started at a specific time
that can be configured.
Volume
The sampling program is started when a certain
totalized flow is reached.
External start
The sampling program is started by a pulse at the
configured binary input.
External duration
The sampling program is active as long as the
configured input has the corresponding level.
Fieldbus(optional):
The sampling program is started by a signal from the
control system.
If Start condition Date/time is selected:
Start date 01.01.2000 to 31.12.2099
Factory setting
DD.MM.YYYY
Set the start date for the sampling program. The
format depends on the option configured under
general settings.
Start time 00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is started.
The format depends on the option configured under
general settings.