Operation Liquistation CSF48
132 Endress+Hauser
Stop condition
(not for External start)
Options
•Program end
• Continuous
•Date/time
•External signal
• Fieldbus (optional)
Factory setting
Program end
Program end:
The device stops sampling automatically once it has
run through the set program.
Continuous:
The device runs through the set program
continuously in an infinite loop. Do not forget to
empty the bottles.
Date/time:
The device stops the set program at a specific time.
External signal:
The device stops the set program if a pulse is sent to a
binary input configured accordingly.
Fieldbus:
The sampling program is stopped by a signal from the
control system.
If Stop condition Date/time is selected:
Stop date 01.01.2000 to 31.12.2099
Factory setting
DD.MM.YYYY
Set the stop date for the sampling program. The
format depends on the option configured under
general settings.
Stop time 00:00:00 to 23:59:59
Factory setting
HH:MM:SS (24h)
Set the time when the sampling program is stopped.
The format depends on the option configured under
general settings.
If Stop condition External signal is selected:
Stop signal input Options
• No program stop input
configured
• Binary input S:x
Factory setting
No program stop input
configured
Select the program stop input.
The binary input must be configured for this function.
Only the inputs configured as a program stop input
are displayed.
If Stop condition PROFIBUS DP or Modbus is selected:
Stop signal input Options
•None
• %0V DO 01
• %0V DO 02
• %0V DO 03
• %0V DO 04
• %0V DO 05
• %0V DO 06
• %0V DO 07
• %0V DO 08
Factory setting
None
Select the program stop input.
Setup subprogram
New
Path: Menu/Setup/Sampling programs
Function Options Info