Operation Liquistation CSF48
134 Endress+Hauser
Example:
- Bottle: 1
- Bottle: 2
...
- Delta (=waiting time):
01:00:00
- Delta (=waiting time):
00:10:00
- Volume: 100 ml
- Volume: 100 ml
1st sampling one hour after program start: 100 ml in bottle 1
2nd sampling 10 minutes later: 100 ml in bottle 2
The sampling table indicates: After the defined "Delta time" (column 2) the volume of column 3 will be dosed into
the bottle of column 1.
If Sampling mode External signal is selected:
Sampling signal input Options
• No sampling input
configured
• Binary input S:x
Factory setting
No sampling input configured
Select the input for the sampling signal.
The binary input must be configured for this function.
Only the configured inputs are displayed.
If Sampling mode PROFIBUS DP or Modbus is selected:
Sampling signal input Options
•None
• %0V DO 01
• %0V DO 02
• %0V DO 03
• %0V DO 04
• %0V DO 05
• %0V DO 06
• %0V DO 07
• %0V DO 08
Factory setting
None
Select the input for the sampling signal. The fieldbus
must be configured for this function.
Change signal
This item only appears if
bottle changeover is set
to fieldbus.
Options
•None
• %0V DO 01
• %0V DO 02
• %0V DO 03
• %0V DO 04
• %0V DO 05
• %0V DO 06
• %0V DO 07
• %0V DO 08
Factory setting
None
Select the input for the bottle change signal.
Path: Menu/Setup/Sampling programs
Function Options Info