Liquistation CSF48 Diagnostics and troubleshooting
Endress+Hauser 203
11.9.3 Program logbook
The following table shows an overview of the exported program logbook and explains the most
important terms in the program logbook.
Entry Example Info
Timestamp 05.05.2010 12:40 Time stamp - the start time in the case of sampling
Event BasicPrgStart Power on
--> Time the device was started
Power failure
--> Time the power failed (to the minute)
BasicPrgStart, StdPrgStart
--> Time the program was started
BasicSampling, StdSampling
--> Entry made during sampling
PrgPartStart, PrgPartStop
--> Time a subprogram is enabled and disabled
PrgStop
--> Time the program was ended
Name Program1 In the case of BasicPrgStart, StdPrgStart, BasicSampling or
PrgStop
--> the name of the program appears
In the case of StdSampling, PrgPartStart or PrgPartStop
--> the name of the subprogram appears
Bottle configuration 12x+6x - PE/glass plate
distribution
The selected bottle configuration is displayed
Left bottle volume 1000 The bottle volume is displayed
--> "Right bottle volume" remains empty for bottle
configurations with different volumes
Right bottle volume 3000
Sampling mode Time-paced CTCV Time-paced CTCV --> in proportion to time
Flow-paced VTCV --> in proportion to volume
Time/flow-paced CTVV --> in proportion to flow
Single sample --> single sample
Sample table --> single sample
--> the sampling mode is displayed
Sampling interval/unit 10 min --> The interval and unit are displayed
Samples/bottle 4 With bottle change
--> Number of samples per bottle
Bottles/sample 0 Multiple bottles
Sampling volume/unit 100 ml Sample volume when sampling