Liquistation CSF48 Diagnostics and troubleshooting
Endress+Hauser 205
11.9.4 Bottle statistics
In "Menu/Diagnostics/Logbooks/Logbook program", select the "Show summary of current
program" item to display the bottle statistics for the sampler. The statistics are displayed for each
individual bottle when the program is started. This gives you detailed feedback on the last
sampling operations.
The statistics are deleted when the following event occurs:
• Program is started
The statistics are selectively overwritten when the following event occurs:
• When the 1st bottle is reached in situations where "Continuous operation" is configured
as the end of the program in the program settings.
The statistics are displayed as follows:
Column Display Info
1 hh:mm The time the first sample was transferred to the bottle is displayed.
2 bt The bottle number is displayed.
3 Smp Displays how often sampling was triggered per bottle.
4 n.s. Indicates the number of times a sample was not taken even though sampling was triggered. This
can occur if the maximum permissible fill volume for the bottle has been reached but the system
is still supposed to transfer samples to the bottle. The "Overfill sensor" message is displayed
while the program is active.
5 n.f. The value indicates how often sampling was canceled since the system was unable to take in any
medium, or enough medium, into the dosing chamber to cover the LF1 probe.
6 ml The sampling volume collected per bottle is displayed.
7 Q The total flow for every bottle is displayed (if connected).