Liquistation CSF48 Diagnostics and troubleshooting
Endress+Hauser 213
Medium LF1 Read only • Medium detection LF1 switchoff
• Medium detection LF2 disconnection from protective
circuit
-> Both "No" at the start
-> If "Yes", clean LF2
Medium LF2 Read only
Inline sampling
(only for version with
sampling assembly)
Action
Sampling activated, to stop
press ESC
Read only
Progress Read only
Distribution arm Action Only for bottle configurations with more than one
bottle.
Test distribution arm Read only When the menu item is activated, the distribution arm
undergoes a test run. Afterwards, the system moves to
each position in succession and the position is displayed.
In the case of plate distribution, the arm moves left and
right to ensure the bottles are numbered consecutively.
Calibrate the distribution arm if the arm is not
positioned precisely over the bottles.
Position Read only
Device reset Options
•OK
•ESC
Restart and keep all the settings
Factory default Options
•OK
•ESC
Restart with factory settings
Settings that have not been saved are lost.
Power supply Read only
• Digital Supply 1: 1.2V
• Digital Supply 2: 3.3V
• Analog Supply: 12.5V
• Sensor Supply: 24V
•Temperature
Detailed list of power supply to instrument.
The actual values can vary without a malfunction
having occurred.
Diagnostics/Systemtest/Reset
Function Options Info